Gas analyzer Vario Plus Industriall

Device Type: Portable

Measurements: O2 / CO / NO / NO2 / SO2 / H2S / CO2 / CH4 / C3H8 / H2 / Temperature / Diff. Pressure / m / s

With the possibility of operation in a semi-stationary mode

• The most accurate gas analyzer in its class!

• Ideal device for ecologists and technologists.

• Number of sensors: 2 to 9 - O2, CO, NO, NO2, SO2, H2S, H2, CH4 or C3H8, temperature, pressure / discharge, differential pressure, + CO2 calculation, alpha, efficiency, flow rate, Mass emissions.

• Up to 6 electrochemical sensors (O2, CO, NO, NO2, SO2, H2S).

• Up to 3 infrared NDIR sensors (O2, CO, CH4 or C3H8).

• Graphic display with backlight (Russian +14 languages) with "ZOOM" function, built-in high-speed printer, gas cooler (Peltier) with automatic condensate removal, memory for 8500 blocks, outputs to PC. Included: built-in batteries, mains adapter, rugged housing, modular gas sampling probe, spare filters, leather trunk.

The universal gas analyzer MRU Vario Plus Industrial is a multicomponent measuring modular system.

This means that this analyzer can be used as a mobile with autonomous power supply, as an accurate instrument for long-term measurements. If necessary, it is possible to use the MRU Vario Plus Industrial in a semi-stationary automatic measurement mode. It is intended for conducting routine tests, measuring Mass emissions, especially recommended for measurements in gas turbines.

Advantages:

• Use of infrared (NDIR) and electrochemical sensors

• Measurement of 9 gases simultaneously (3 infrared + up to 6 electrochemical)

• Graphic display with backlight (Russian +14 languages) with "ZOOM" function

• Operation in automatic measurement mode

• Built-in gas cooler (Peltier) with automatic condensate removal

Gas temperature measurement up to 1700 ° C

• 8 programmable analog outputs 4 ... 20mA

• Ability to work with heated gas lines

• Allows you to work with the remote control

• Long service life

• Designed for active use

• Pressure / discharge / differential pressure measurement

• High-speed printer without carriage (print time 3 sec)

• Very robust case with carrying strap

• Suitable for all types of burners

• Easy to use

• Different lengths and material of probes

• Russian version of the on-screen menu of the device

• Built-in memory for 8500 measurements

• RS 232 interface for data transfer to a PC


Specifications:

Electrochemical sensors

Measured values

Limits of permissible errors

O2

Measurement range:

0 - 21,0 %

Measurement error:

± 0,2 % абс.

CO

(с H2компенсацией)

Measurement range:

0 - 4.000 ppm, Overload to 10.000 ppm

Measurement error:

± 20 ppm or 5 % from m.value / > 4.000 ppm

10 % from m.value / > 4.000 ppm

CO very high (option)

Measurement range:

0 - 40.000 ppm, overload to 100.000 ppm

Measurement error:

± 200 ppm or 5 % from m.value / > 40.000 ppm

10 % from m.value / > 40.000 ppm

NO (option)

Measurement range:

0 - 1.000 ppm, overload to 5.000 ppm

Measurement error:

± 5 ppm or 5 % from m.value / > 1.000 ppm

10 % from m.value / > 1.000 ppm

NO2 (option)

Measurement range:

0 - 200 ppm, перегрузка до 1.000 ppm

Measurement error:

± 5 ppm or 5 % from m.value / > 200 ppm

10 % from m.value / > 200 ppm

SO2 (option)

Measurement range:

0 - 2.000 ppm, overload to 5.000 ppm

Measurement error:

± 10 ppm or 5 % from m.value / > 2000 ppm

H2S (option)

Measurement range:

0 - 50 ppm, overload to 500 ppm

Measurement error:

± 5 ppm or 5 % from m.value / < 50 ppm

H2 (option)

Measurement range:

0 - 1%, overload to 2%

Measurement error:

± 0,02% or 5 % from m.value / < 1%

10 % from m.value / > 1%

With multi-gas IR module

CO

Measurement range:

min. 0 - 10.000 ppm to max. 0 - 30 %

Measurement error:

± 40 ppm or 5 % from m.value

CO2

Measurement range:

min. 0 - 3 % to max. 0 -30 %

Measurement error:

± 0,50 % or 5 % from m.value

CXHY as C3H8

Measurement range:

min. 0 - 10.000 ppm to max. 0 - 5 % as C3H8

Measurement error:

± 20ppm or 5% from m.value

CXHY as CH4

Measurement range:

min. 0 - 10.000 ppm to max. 0 - 5 % as CH4

Measurement error:

± 60 ppm or 5% from m.value

Flue gas temperature TA

Measurement range:

0 - 650 °C Stainless steel probe

Measurement error:

0 - 1.100 °C Inconel alloy probe

Measurement range:

0 - 1.750 °C with ceramic probe

Measurement error:

± 2 °C / < 200 °C

1 % from m.value > 200 °C

Combustion temperature TL

Measurement range:

0 - 100 °C

Measurement error:

± 1 °C

Draft / depression

Measurement range:

± 100 hPа

Measurement error:

± 0,03 hPа or 1% from m.value

Differential pressure

Measurement range:

± 100 hPа

Measurement error:

± 0,03 hPа or 1% from m.value

Gas flow rate (with Pitot tube)

Measurement range:

1 m/s to 100 м/с

Measurement error:

± 1 m/s or 1 % from m.value

Mass emissions:

g/ s, ...

Estimated values: (Dependent on the type of fuel)

CO2 (Estimated for an unidentified direct measurement of CO2 (calculation is based on the selected fuel type and the measured value of O2)

Measurement range:

0 - CO2 max

Dew point:

°C

Losses with flue gases qA:

0 - 99,9 %

Efficiency ή:

0 - 120 %

Dimension of measurement data:

mg/Нm3, Relative to O2, mg / kWh, NOx as mg / Nm3NO2, mg / s.


Common data:

Operation temperature:

+ 5 - + 45 °C, max. 95 % without condensation

Storage temperature:

- 20 - + 50 °C

Supply voltage:

Net 110 - 230 Vac, 50/60 Hz

12 V / 1,8 Ah with internal battery, 2 hours of operation

Protection class:

IP 21

Weight:

About 7.000 g (without bag and accessories)

Dimensions:

530 x 490 x 310 mm

Our projects