Gas analyzer semistationary MGA5 +
Device type: Semi-stationary (Up to 30 days of continuous measurement)
Measurements: O2 / CO / NO / NO2 / SO2 / H2S / CO2 / CH4 / C3H8 / H2 / Temperature / Diff. Pressure / m / s
High-precision mobile semi-stationary gas analyzer
• With the ability to quickly change ranges of measurement and calibration without gas (CBC) cylinders!
• Number of gases to be measured: 6 - (O2, CO, NOx, converter (NO2 to NO), CH4 or SO2), temperature, pressure / discharge, differential pressure, + CO2 calculation, alpha, efficiency, flow rate, volumetric flow, Mass Emissions.
• Graphic display with backlight (Russian +14 languages) with "ZOOM" function, external high-speed printer, 2 gas refrigerators (Peltier) with dew point stabilization + 5 ° C, with automatic condensate removal, memory for 8500 blocks, outputs to PC.
Precision gas analyzer MRU MGA5+ is a unique semi-stationary measuring system that allows to conduct high-precision technological and ecological official measurements with the accuracy of the best stationary gas analyzers.
It is intended for carrying out of official and responsible tests, measurements, it is especially recommended for carrying out of measurements in gas turbines.
Advantages:
Using high-precision infrared (NDIR) multichannel sensors with the ability to quickly change the measuring ranges (from 200 to 1.000ppm), allows you to conduct the most accurate measurements. In this case, if the measuring range set by the user is exceeded (up to 10 times), the sensor does not enter saturation mode, and the analyzer displays the current value.
In addition, the MRU MGA5 + has a unique automatic calibration system without gas (CGS) cylinders (option).
MGA5 + is a mobile gas analysis system that combines mobility and highest measurement accuracy.
• Graphic display with backlight (Russian +14 languages) with "ZOOM" function
• Operation in automatic measurement mode
• Built-in gas cooler (Peltier) with automatic condensate removal
Gas temperature measurement up to 1700 ° C
• 8 programmable analog outputs 4 ... 20mA
• Ability to work with heated gas lines
• Allows you to work with the remote control
• Long service life
• Designed for active use
• Pressure / discharge / differential pressure measurement
• High-speed printer without carriage (print time 3 sec)
• Very robust case with carrying strap
• Suitable for all types of burners
• Easy to use
• Different lengths and material of probes
• Russian version of the on-screen menu of the device
• Built-in memory for 8500 measurements
• RS 232 interface for data transfer to a PC
Specifications: |
|
Electrochemical sensor O2 (service life 5 years) |
|
Measurement range |
0 - 25,0 % |
Accuracy / Resolution |
± 0,2 % об. / 0,01 % |
Multi-gas IR module |
|
Range of CO |
0-200 ppm to 0-1000 ppm |
Error |
± 2% of the range or 5% of the measurement. Val. |
Resolution |
1 ppm |
Range of CO2 |
0-4% to 20% by volume. |
Error |
± 2% of the range or 5% of the measurement. Val. |
Resolution |
0.0001% vol. |
Range of NO |
0-200 ppm to 0-1000 ppm |
Error |
± 2% of the range or 5% of the measurement. Val. |
Resolution |
1 ppm |
Range CH4 |
0-200 ppm to 0-1000 ppm |
Error |
± 2% of the range or 5% of the measurement. Val. |
Resolution |
1 ppm |
SO2 Range |
0-200 ppm to 0-1000 ppm |
Error |
± 2% of the range or 5% of the measurement. Val. |
Resolution |
1 ppm |
Catalytic converter NO2 to NO |
|
NO2 range |
0-100 ppm |
Т gas temperature |
|
Range |
0 - 650 ° C (stainless steel tube) |
Range |
0 - 1100 ° C (Inconel alloy tube) |
Range |
0 - 1750 ° C (ceramic tube) |
Error |
± 2 ° C to 200 ° C, 1% of the change. St. 200 ° C |
Air temperature of the burner |
|
Range / Error |
0 - 100 °C / ± 1 °C |
Differential pressure (option) |
|
Range |
± 100 hPa |
Error |
± 0.03 hPa or 1% of the change. Val. |
Flow rate |
|
Range |
1 m / s to 100 m / s |
Error |
± 1 m / s or 1% of change. Val. |
Common specifications |
|
Terms of Use |
+ 5 ° C - + 45 ° C |
Storage temperature |
Not more than 95% RH, without condensation |
Main fuse |
- 20 ° C to + 50 ° C |
Speed |
10 amp |
Warm-up time |
20 seconds, without transport delay |
Display |
30 min |
Output signals |
Graphic LCD display with backlight |
Sample preparation |
8 analog outputs 4-20mA, digital interface RS 232 |
Gas test control |
Built-in gas cooler with dew point + 5 ° C, particle filter <1μm |
Software calibration. Calibration gases for each gas channel, clean air for zero setting |
|
Protection class |
IP 21 |
Dimensions |
|
Main Control Unit |
500 x 520 x 295 mm / 17 kg |
IR Analyzer |
500 x 520 x 205 mm / 20 kg |