The Model T200 Chemiluminescence NO/O2/NOx Analyzer
The Model T200 NO / NO2 /NOx analyzer uses the proven chemiluminescence detection principle and advanced electronics to allow accurate, dependable, continuous measurements for ambient air quality, stack gas monitoring and other applications.
Available with NumaView™ premium T Series software
-
Large, vivid, and durable color touchscreen display
-
All other T Series instrument platform features
-
Lifetime technical support by phone and email
-
Standard two-year warranty
T200 Specifications
Ranges |
Min: 0-50 ppb Full scale Max: 0-20,000 ppb Full scale (selectable, dual ranges and auto ranging supported) |
Measurement Units |
ppb, ppm, μg/m3, mg/m3 (selectable) |
Zero Noise |
< 0.2 ppb (RMS) |
Span Noise |
< 0.5% of reading (RMS) above 50 ppb |
Lower Detectable Limit |
0.4 ppb |
Zero Drift |
< 0.5 ppb/24 hours |
Span Drift |
< 0.5% of full scale/24 hours |
Lag Time |
20 seconds |
Rise/Fall Time |
< 60 seconds to 95% |
Linearity |
1% of full scale |
Precision |
0.5% of reading above 50 ppb |
Sample Flow Rate |
500 cc/min ±10% |
Power Requirements |
100V-120V, 220V-240V, 50/60 Hz |
Analog Output Ranges |
10V, 5V, 1V, 0.1V (selectable) |
Recorder Offset |
±10% |
Included I/O |
1 x Ethernet: 10/100Base-T 2 x RS232 (300-115,200 baud) 2 x USB device ports 8 x opto-isolated digital outputs 6 x opto-isolated digital inputs 4 x analog outputs |
Optional I/O |
1 x USB com port 1 x RS485 8 x analog inputs (0-10V, 12-bit) 4 x digital alarm outputs Multidrop RS232 3 x 4 - 20mA current outputs |
Operating Temperature Range |
5 - 40ºC |
Dimensions (HxWxD) |
7” x 17” x 23.5” (178 x 432 x 597 mm) |
Weight |
Analyzer 18 kg External pump: 15 lbs (7 kg) |
Certifications |
MCerts: Sira MC0500067/07 US EPA: EQSA-0495-100 Russian standard: 50500-12 |
* All certifications apply for legacy or NumaView™ T Series analyzer software
NumaView™ software is available as a no-charge option that must be specified at the time of purchase.
Specifications subject to change without notice.
All specifications are based on constant conditions.
Our projects
Useful
Reviews
Head count
-
Как Вам наш сайт ENGLISH?
Нравиться 0 (%) Не нравиться 0 (%) Воздержусь 0 (%)